“PNT Digital
with Inductotrace option ”
- respiratory inductive plethysmographe RIP
Respiratory Monitoring

Non-Invasive
Comfortable to wear
Quantitative Tidal Volume
Breathing Patterns
Adult
- Pediatric - Neonate
Principle of operation
Ø
The “Digital Inductotrace” is a respiratory inductive
plethysmographe RIP, designed for non invasive respiratory monitoring which
brings the combination of accuracy (± 10% of spirometry), reliability, and
versatility in a lot of physiological measurement applications.
Ø
Two elastic bands
with insulated wires called “Inductobands” wrapped around the chest and
abdomen provide signals through frequency oscillators to a Digital
Calibrator/Demodulator Unit.
Any changes in cross-sectional area of thoracic and abdominal regions (Inspiration-Expiration)
provide a proportional and linear change in coil inductance from
“Inductobands” and measure ribcage and abdomen compartmental volume
excursions in amplitude and phase.
Ø
Unlike other devices for measuring chest and abdominal wall movement, can
be calibrated with a pneumotachometer or a spirometer to provide not only the
pattern of breathing but actual tidal volumes and phase regardless of position
changes of the patient.
Features
·
The patient can
breathe freely and naturally, this eliminates artifacts and measurement errors
introduced by the interaction of mouthpieces or masks.
(interaction on respiration level,
tidal volume, hyperventilation, etc…)
·
Less discomfort for the patient which means long term monitoring
·
Large alpha numerical and graphical display
·
High sensitivity, stability and linearity over the full range of
inductive plethysmography
·
Selection of AC or DC coupling and full control of Inductive
Plethysmography gains
-
Sampling rate at 250 pts/sec
(high definition of onset of mechanical inspiration)
-
4 Analog output of Inductive
Plethysmography waveforms for polygraph interface +/- 1V, 5V ,10V DC for
Vth, Vab, Vsum and Fsum when used as a Respiratory Inductive Plethysmograph
(RIP) or integrated devices spirometer, capnometer etc…
-
4 Analog inputs for
temporary or continuous simultaneous display and recording of signals from
other devices
Applications
(See
Bibliography)
·
Non-Invasive Ventilation in acute Respiratory Failure
·
Apnea/Hypopnea and Upper Airway Flow Limitation
·
Respiratory Drive and Breathlessness (Dyspnea)
·
Depression of Respiratory Drive in Pain Management
·
Respiratory rehabilitation
·
Hemi thoracic Inductive Plethysmography
·
Substance Abuse
·
Sleep Disorders (detection of obstructive or central apnea, hypopnea,
Cheyne-Stokes, etc…)
·
Respiratory Efforts and Swallows from Neck Inductive Plethysmograph
Ø
Respiratory Mechanics
·
Forced expiratory volume
Speech breathing
Fig. 1 & 2
- Breathing
Abnormalities During Steep.
Martin Tobin, MB, MRCPI,- Martin Cohn, MD; Marvin Sackner, MD; Arch lntem, Med.;
Vol 143, June 1983.

Fig 1.
- Central sleep apnea. There is absence of abdominal (ABD), rib cage
(RC), and sum (VT) movements during central apnea associated with small fall of
arterial oxygen saturation (02Sat) measured via ear oximetry.
Fig
2.
- Obstructive sleep apnea.
This depicts obstructive apnea terminated by deep breaths and another
obstructive apnea terminated by three deep breaths. Apnea in midportion of
recording is marked by
absence of sum movements (VT) despite respiratory efforts indicated by paradoxic
movement of rib cage (RC) (movement downward) and abdominal (ABD) (movement
upward) compartments that is associated with marked fall in arterial oxygen
saturation (02Sat) measured by ear oximetry.
Options and upgrades for PNT Digital
Inductotrace
Optional integrated pneumotachometer or spirometer is available for
calibration of respiratory scale factors
Optional integrated pulmonary pressure for arterial blood pressure
monitoring on request
Optional integrated capnometer for respiratory monitoring on request.
Optional integrated pulse oximeter for respiratory monitoring on request
Optional integrated non invasive blood pressure for arterial blood
pressure monitoring on request
Optional integrated digital outputs USB , RS 232, I²C for further
computer processing on request.
Optional integrated analog inputs from external equipment on request.
Optional special configuration integrated or not integrated on request.
Order information:
|
Ref.
No. |
Description |
Size |
|
10.5005 |
Inductoband |
PREMATURE - torso girth 17.78-22.86 cm (7" to 9") |
|
10.5010 |
Inductoband |
EXTRA
SMALL - torso girth 25.40-38.10 cm (10" to 15") |
|
10.5012 |
Inductoband |
INFANT/BABY - torso girth 30.48-43.18 cm (12" to 17") |
|
10.5020 |
Inductoband |
TODDLER - torso girth 35.56-50.80 cm (14" to 20") |
|
10.5025 |
Inductoband |
CHILD/TEEN - torso girth 45.72-63.50 cm (18" to 25") |
|
10.5028 |
Inductoband |
SMALL
ADULT - torso girth 55.88-71.12 cm (22" to 28") |
|
10.5032 |
Inductoband |
MEDIUM
ADULT - torso girth 66.04-81.28 cm (26" to 32") |
|
10.5036 |
Inductoband |
REGULAR
ADULT - torso girth 71.12-91.44 cm (28" to 36") |
|
10.5040 |
Inductoband |
LARGE
ADULT - torso girth 81.28-101.60 cm (32" to 40") |
|
10.5048 |
Inductoband |
JUMBO
ADULT - torso girth 86.36-121.92 cm (34" to 48") |
|
10.5054 |
Inductoband |
REGAL
ADULT - torso girth 106.68-137.16 cm (42" to 54") |
|
10.5066 |
Inductoband |
X-LARGE
ADULT - torso girth 121.92-167.64 cm (48" to 66") |
|
|
|
|
|
10.5100L |
Cable
for Inductolock |
|
|
10.5005L |
Inductolock |
PREMATURE - torso girth 17.78-22.86 cm (7" to 9") |
|
10.5010L |
Inductolock |
EXTRA
SMALL - torso girth 25.40-38.10 cm (10" to 15") |
|
10.5012L |
Inductolock |
INFANT/BABY - torso girth 30.48-43.18 cm (12" to 17") |
|
10.5020L |
Inductolock |
TODDLER - torso girth 35.56-50.80 cm (14" to 20") |
|
10.5025L |
Inductolock |
CHILD/TEEN - torso girth 45.72-63.50 cm (18" to 25") |
|
10.5028L |
Inductolock |
SMALL
ADULT - torso girth 55.88-71.12 cm (22" to 28") |
|
10.5032L |
Inductolock |
MEDIUM
ADULT - torso girth 66.04-81.28 cm (26" to 32") |
|
10.5036L |
Inductolock |
REGULAR
ADULT - torso girth 71.12-91.44 cm (28" to 36") |
|
10.5040L |
Inductolock |
LARGE
ADULT - torso girth 81.28-101.60 cm (32" to 40") |
|
10.5048L |
Inductolock |
JUMBO
ADULT - torso girth 86.36-121.92 cm (34" to 48") |
|
10.5054L |
Inductolock |
REGAL
ADULT - torso girth 106.68-137.16 cm (42" to 54") |
|
10.5066L |
Inductolock |
X-LARGE
ADULT - torso girth 121.92-167.64 cm (48" to 66") |
|
|
|
|
|
10.5220R |
Small Retainer for Inductoband 10.5005, 10.5010, 10.5012 |
|
10.5228R |
Child Retainer for Inductoband 10.5020, 10.5025 |
|
10.5236R |
Adult Retainer for Inductoband 10.5028, 10.5032, 10.5036 |
|
10.5248R |
Large Retainer for Inductoband 10.5040, 10.5048 |
|
10.5266R |
X-Large
Retainer for Inductoband 10.5054, 10.5066 |
|
|
|
|
|
Accessories for Digital Inductotrace |
|
104026 |
Spirobag for calibration procedure |
|
106501 |
Cables
for recorder connections ‘4 pieces) |
|
106500 |
Respitrace cable with 2 transducer -oscillators |
|
106502 |
Respiband connecting wires (Red/green) with mini connector |
|
|
|
|
|
|
| |
|
|
|
|
Mechanical |
|
|
Size |
Size: (H 17.5cmx W 29cm x D 26cm) (6.8"H x 11.4"W x 10.2"D)
|
|
Weight |
Weight: 2.2lbs (1.0kg) |
|
Electrical: |
|
|
Power |
External Power Supply: 100-240VAC, 50/60Hz / T 0.5 A |
|
Protection Class |
Class I |
|
Degree of protection against electric chock: |
B |
|
Mode of operation: |
Intermittent |
|
Environmental: |
|
|
Operating temperature |
10-40oC Humidity: 20% to 70% (non condensing) |
|
Explosive area: |
This equipment must not be used in an explosive or combustible
situation |
|
Altitude operation |
0 - 3000 m |
|
Warm-up time: |
30 sec |
|
Integrated Microprocessor |
|
|
Processor |
Cygnal and Cypress |
|
Display: |
|
|
LCD size |
320x240 B/W LCD display |
|
Sensors |
|
|
Oscillators Transducers |
Signal Oscillator between 100khz – 800khz depending the
respiband |
|
Inductobands Sensors |
Inductobands pecial sizes for adult , infant and neonate
|
|
|
Parameters: |
|
|
Thorax volume Vth |
Range: See inductoband product table
Linearity: <1% |
|
Abdominal volume Vab |
Range: See inductoband product table
Linearity: <1% |
|
Total volume = Vth + Vab |
Range: See inductoband product table
Linearity: <1% |
|
Total flow |
Range: See inductoband product table
Linearity: <1% |
|
Electrical outputs |
|
|
4 channels Analog Output: |
analog output Voltage -/+ 1V, 5V, 10V - Low impedance :
<10 Ohm |
|
Resolution |
0.5mV (12 bit dynamic range), |
|
connections |
BNC output connectors and BNC terminations on provided cable
+/-2m |
|
Certification/Safety standards: |
|
|
93/42/EEC |
Medical Device Directive |
|
EN60601-1 |
General Requirements for Safety |
|
EN60601-1-1 |
Safety Requirements for Medical Electrical Systems |
|
EN60601-1-2 |
Electromagnetic compatibility |
|
CE approval: |
|
|
|
Final inspection and test of cardiopulmonary function test
equipment in class I
(93/42/EEC Annex VI Section 3.2 - Product quality assurance)
|
|
|